SD-X01 Specifications
Not all of the JMS products shown on this website are registered in, approved for sale in, or available for purchase in, all countries or regions where this website is accessible. Please consult with your local JMS representative,
distribution company or customer service center for additional information.
Not all of the following products may be classified as medical devices. Please note that the definition of medical devices differs in each country.
Water supply | Supply pressure | 39.2 to 196.1 kPa (0.4 to 2.0 kgf/cm2) |
---|---|---|
Supply temperature | 10 to 30°C (when the 1.0 kW heater is used) To be lower than the dialysate temperature setting value by 3°C |
|
Dialysate temperature | Setting range | 33.0 to 39.9°C |
Display range | 0.0 to 50.0°C | |
Dialysate conductivity | Measurement range | 12.0 to 18.0 mS/cm |
Dialysate flow rate | Setting range | 300 to 700mL/min |
Ultrafiltration rate | Setting range | 0.00 to 5.99 L/h |
UF goal | Setting range | 0.00 to 39.9 L |
Dialysis time | Setting range | 00:00 to 99:59 |
Ultrafiltration time | ||
Venous pressure sensor | Display range | -500 to +700 mmHg |
Pressure of blood inlet (PBI) sensor (optional) | ||
Pressure of dialysate outlet (PDO) sensor | ||
Blood pump | Flow rate setting range | Tube L: 30 to 500 mL/min |
Tube S: 30 to 500 mL/min | ||
Syringe pump | Flow rate setting range | 0, 0.1 to 10.0 mL/h |
Syringe used | 20/10 mL syringe | |
Blood leak detector | Method | Optical |
Detection sensitivity | Blood 300 ppm (Hematocrit 32%, 37°C) | Air detector | Air detection method | Ultrasonic method |
Individual air bubble detection sensitivity | 0.07 mL or more (at a blood flow rate of 200 mL/min) | |
Microbubble detection sensitivity | 0.3 μL or more (at a blood flow rate of 250 mL/min) | |
Substitution pump(optional) | Substitution flow rate setting range | 1.8 to 30.00 L/h |
Rinse/disinfection | Acetic acid (10-30%) | |
Sodium hypochlorite (4-6%) | Electric ratings | Rated voltage | 100 to 120 VAC (100 V system) |
220 to 240 VAC (200 V system) | Frequency | 50/60 Hz | Power consumption | 1.5 kVA (1.0 kW for heater) |
1.25 kVA (0.75 kW for heater) (optional) | ||
1.0 kVA (0.5 kW for heater) (optional) | ||
Protection against water penetration | IPX1 | |
Dimensions | 350 mm (W) × 580 mm (D) × 1410 mm (H) * Excluding protrusions |
|
Weight | 80 kg |