JMS logo

Japanese

JMS logo

  1. HOME
  2. Products
  3. Dialysis
  4. Single Patient Dialysis Machine

SD-X01 Specifications

Not all of the JMS products shown on this website are registered in, approved for sale in, or available for purchase in, all countries or regions where this website is accessible. Please consult with your local JMS representative, distribution company or customer service center for additional information.
Not all of the following products may be classified as medical devices. Please note that the definition of medical devices differs in each country.

Water supply Supply pressure 39.2 to 196.1 kPa (0.4 to 2.0 kgf/cm2)
Supply temperature 10 to 30°C (when the 1.0 kW heater is used)
To be lower than the dialysate temperature setting value by 3°C
Dialysate temperature Setting range 33.0 to 39.9°C
Display range 0.0 to 50.0°C
Dialysate conductivity Measurement range 12.0 to 18.0 mS/cm
Dialysate flow rate Setting range 300 to 700mL/min
Ultrafiltration rate Setting range 0.00 to 5.99 L/h
UF goal Setting range 0.00 to 39.9 L
Dialysis time Setting range 00:00 to 99:59
Ultrafiltration time
Venous pressure sensor Display range -500 to +700 mmHg
Pressure of blood inlet (PBI) sensor (optional)
Pressure of dialysate outlet (PDO) sensor
Blood pump Flow rate setting range Tube L: 30 to 500 mL/min
Tube S: 30 to 500 mL/min
Syringe pump Flow rate setting range 0, 0.1 to 10.0 mL/h
Syringe used 20/10 mL syringe
Blood leak detector Method Optical
Detection sensitivity Blood 300 ppm (Hematocrit 32%, 37°C)
Air detector Air detection method Ultrasonic method
Individual air bubble detection sensitivity 0.07 mL or more (at a blood flow rate of 200 mL/min)
Microbubble detection sensitivity 0.3 μL or more (at a blood flow rate of 250 mL/min)
Substitution pump(optional) Substitution flow rate setting range 1.8 to 30.00 L/h
Rinse/disinfection Acetic acid (10-30%)
Sodium hypochlorite (4-6%)
Electric ratings Rated voltage 100 to 120 VAC (100 V system)
220 to 240 VAC (200 V system)
Frequency 50/60 Hz
Power consumption 1.5 kVA (1.0 kW for heater)
1.25 kVA (0.75 kW for heater) (optional)
1.0 kVA (0.5 kW for heater) (optional)
Protection against water penetration IPX1
Dimensions 350 mm (W) × 580 mm (D) × 1410 mm (H)
* Excluding protrusions
Weight 80 kg